ProductsProducts / Evidence
Recalled and not-approved radon monitors
Short answer
Health Canada has issued five recalls and two advisories since 2022 for consumer radon monitors that measure radon inaccurately, and Canada's radon proficiency program (C-NRPP) lists 14 consumer monitors it does not approve. If you own one, stop using it and test again with a lab-analyzed kit or a monitor with an independent result.
Key takeaways
- Canada's radon proficiency program, C-NRPP, lists 14 consumer radon monitors as not approved. Its 2024/25 report says plainly that C-NRPP does not recommend them.
- Health Canada has published seven notices about consumer radon monitors since 2022: five recalls and two advisories. Each found the devices measure radon inaccurately; two said they report a low value regardless of the actual radon level.
- If you own one, Health Canada's advice is to stop using it and dispose of it as electronic waste. Then test again with a device you can trust.
- Every device here was sold on Amazon.ca. We could not confirm on 1 October 2026 whether any of them is still listed in the United States, so check the model number before you buy any monitor.
- A monitor that is missing from both lists has not failed anything. Health Canada says a device may be on neither list because it is still being tested or reviewed.
Which radon monitors have been recalled or are not approved
Two Canadian bodies publish what is, as far as we know, the most complete public record of consumer radon monitors that do not measure radon reliably. The Canadian National Radon Proficiency Program (C-NRPP) tests consumer monitors so that Canadians can compare them on performance alone. Health Canada tests new models at its lab in Ottawa and issues recalls when a device fails. C-NRPP has found that many recent arrivals on the market performed so poorly that they have since been recalled.
This page lists every device on C-NRPP's not-approved list and every Health Canada notice that list points to, with the notice number, its date, the model numbers and what Health Canada found. It names no device to buy. For the monitors that passed C-NRPP's testing, and the many that were never entered, see radon detectors and monitors and independent accuracy results explained.
Calculated by Radon411
14
consumer monitors on C-NRPP's not-approved list, counted from its web page
Calculated by Radon411
7
Health Canada notices about consumer radon monitors that C-NRPP's list links to: 5 recalls and 2 advisories
Every device on C-NRPP's not-approved list
C-NRPP's consumer monitor list has a "NOT APPROVED" section, and marks the devices Health Canada has acted on. The table below lists each device exactly as C-NRPP names it, in A to Z order, and shows which Health Canada notice actually names that product. Where C-NRPP points to a notice about a different product, the last column says so.
| Device, as C-NRPP lists it | C-NRPP status | Health Canada notice that names it | Where C-NRPP points, if different |
|---|---|---|---|
| Air Steward | Not approved | Health Canada recall RA-64544, 2022: the Air Steward Portable Radon Monitor | Same notice |
| Bootu RN-80 | Not approved | None. C-NRPP lists no Health Canada notice for it. | C-NRPP points to no notice |
| Boyd Gresham Radon Detector | Not approved | Health Canada advisory RA-77383, 2025: the Boyd Gresham "Toolhome" Radon Detector, model HRDM-01B | Same notice |
| CRADTEC PRM-02H | Not approved | None. C-NRPP lists no Health Canada notice for it. | C-NRPP points to no notice |
| CRADTEC PRM-03H | Not approved | None. C-NRPP lists no Health Canada notice for it. | C-NRPP points to no notice |
| Funny Kitchen HRDM-02 | Not approved | Health Canada advisory RA-77383, 2025: the Funny Kitchen Home Radon Detector, model HRDM-02 | C-NRPP points this row to the Hanchen recall instead. |
| HAKINAKU Smart Radon Gas Detector | Not approved | None. C-NRPP lists no Health Canada notice for it. | C-NRPP points to no notice |
| Hanchen Home Radon Detector | Not approved | Health Canada recall RA-77403, 2025 | Same notice |
| INKBIRD Home Radon Meter | Not approved | None of the seven notices names this model. | C-NRPP marks it recalled and points to the Life Basis recall, which names models LSRM001 and LCARM001. |
| INKBIRD INK-RD2 | Not approved | Health Canada recall RA-77461, 2025: the InkBird Wifi Radon Detector model RD-2, labeled "Ink-RD2" | Same notice |
| LifeBasis INK-RD2 | Not approved | None of the seven notices names a Life Basis INK-RD2. | C-NRPP points to the Life Basis recall, which names models LSRM001 and LCARM001. C-NRPP's 2024/25 report lists "LifeBasis RN-55" in this place. |
| LifeBasis LCARM001 | Not approved | Health Canada recall RA-77444, 2025: the Life Basis Smart Radon Detector, LCARM001 | Same notice |
| Radon Guard | Not approved | None of the seven notices names Radon Guard. | C-NRPP points to the Elifecity advisory for it, in its 2023 report as well as now. |
| Spolehli Radon Detector | Not approved | Health Canada recall RA-77362, 2025 | Same notice |
The seven Health Canada notices, one by one
Health Canada published two kinds of notice here. Five are a consumer product recall, which reports the company's own figures for units sold. Two are a consumer product advisory: in both, Health Canada says it contacted the foreign selling company and that the product had been removed from sale on Amazon.ca. Both kinds tell owners to stop using the device. The table gives each notice's identification number and original published date.
| Notice | Type and date | Products and model numbers | What Health Canada found | Sold in Canada |
|---|---|---|---|---|
| RA-64529: Elifecity Portable Radon Meter | Advisory, 2022-09-07 | Elifecity Portable Radon Meter | It reported a low value whatever the real radon level, with errors greater than 90%. Removed from sale on amazon.ca. | Amazon.ca and potentially other online websites; no unit count given |
| RA-64544: Air Steward Portable Radon Monitor | Recall, 2022-09-13 | Air Steward Portable Radon Monitor | It may report a low value regardless of the actual radon level, with errors greater than 90%. | About 100 units, 2018 to 2020, on Amazon.ca |
| RA-77362: Spolehli Radon Detector | Recall, 2025-05-01 | Spolehli Radon Detector, Amazon.ca ASIN B0BXSKZJZG | Health Canada's Radiation Protection Bureau found it measures inaccurate radon levels. | 9 units, June to December 2024, only on Amazon.ca |
| RA-77383: various radon gas detectors sold on Amazon.ca | Advisory, 2025-05-07 | Funny Kitchen Home Radon Detector HRDM-02 (ASIN B08RZ568XL); Boyd Gresham "Toolhome" Radon Detector HRDM-01B (ASIN B0CYGHXK7G) | The Radiation Protection Bureau found they measure inaccurate radon levels. Removed from sale on Amazon.ca. | 46 Funny Kitchen units, November 2021 to July 2024; 2 Boyd Gresham units, June 2024 |
| RA-77403: Hanchen Home Radon Detector | Recall, 2025-05-15 | Hanchen Home Radon Detector, ASIN B0C9MN1NVZ | The Radiation Protection Bureau found it measures inaccurate radon levels. | 24 units, November 2023 to June 2024, only on Amazon.ca |
| RA-77444: various Life Basis radon detectors | Recall, 2025-05-30 | Life Basis Radon Detector LSRM001 (ASIN B0B41M8X3W); Life Basis Smart Radon Detector LCARM001 (ASIN B0CLRMJ7KW) | The Radiation Protection Bureau found they measure inaccurate radon levels. | 45 LSRM001 and 79 LCARM001 units |
| RA-77461: InkBird Wifi Radon Detector model RD-2 | Recall, 2025-05-30 | InkBird Wifi Radon Detector RD-2, labeled "Ink-RD2", ASIN B0C3B63VD9 | The Radiation Protection Bureau found it measures inaccurate radon levels. Owners can contact InkBird for a full refund. | 52 units, September 2023 to April 2025 |
The five 2025 notices share one finding: sampling and evaluation by Health Canada's Radiation Protection Bureau determined that the product measures inaccurate radon levels and likely poses a danger to human health or safety. The two 2022 notices are more specific. Each device could report a low value whatever the real radon level was, which is the failure that matters most, because a falsely low reading tells a household it has no problem.
“This device may report a low value regardless of the actual radon level, with errors greater than 90%.”
Recalled, not approved and not tested are three different things
- Recalled (Health Canada)
- Health Canada found the device measures radon inaccurately and published a recall or advisory. Recalls carry a legal rule in Canada: the Canada Consumer Product Safety Act prohibits recalled products from being redistributed, sold or even given away there.
- Not approved (C-NRPP)
- The device appears on C-NRPP's not-approved list. Some of these were recalled; four (the Bootu RN-80, both CRADTEC models and the HAKINAKU detector) carry no Health Canada notice on C-NRPP's list.
- Not included (C-NRPP)
- C-NRPP has published no result for the device either way. Manufacturers sign up to enter C-NRPP's annual comparison, so many monitors sold in the United States have simply never been entered. That is an absence of evidence, not a failed test.
Health Canada draws the same line. It and C-NRPP watch the market and test electronic radon monitors, and devices that fail the minimum performance requirements are recalled. A device can be missing from both lists simply because it is still being tested, or is under review or regulatory action.
| Status | Who decides | What it means for a reading | Where we show it |
|---|---|---|---|
| Approved in C-NRPP's 2024/25 evaluation | C-NRPP, after four rounds in a reference radon chamber | The model performed acceptably in that round of testing on Canadian units | Radon detectors and monitors |
| Not included in C-NRPP's evaluation | Nobody: the maker did not enter it, or C-NRPP has not published a result | No independent result either way | Radon detectors and monitors, with the maker's own claims attributed |
| Not approved by C-NRPP | C-NRPP | Do not rely on its readings | This page only. We publish no product page for it. |
| Recalled or subject to an advisory | Health Canada | Stop using it | This page only. We publish no product page for it. |
Are these monitors still sold in the United States?
Every Health Canada notice above concerns units sold on Amazon.ca. On 1 October 2026 we tried to check Amazon.com for each model on this page and for each Amazon.ca listing number Health Canada published. Amazon.com answered almost every automated check with a security page instead of results, so we could not confirm whether any of these models is listed for sale in the United States today. We will not guess. Until we can check through Amazon's own product data, we treat every model here as one you might still find for sale.
Find the model number
It is usually on the box, the label on the back of the device or the manual. The InkBird recall, for example, says its "Ink-RD2" label is on the packaging.
Compare it with both tables on this page
Match the brand and the model number, not just the look of the device. Several recalled monitors are small black or blue boxes with "Radon Detector" printed under the screen.
Check the listing's identification number
Health Canada published the Amazon.ca listing number (ASIN) for most recalled models. If a listing you are looking at shows one of those numbers, treat it as the recalled product.
Look for an independent result
Before you buy any monitor, see whether it appears in C-NRPP's latest results. If it does not, the maker's accuracy claims are the only evidence, and our hub says so beside each one.
If you own one of these monitors
Stop using it
Every notice says the same thing first: stop using the device at once. Do not pass it on.
Ask the seller for a refund
The InkBird notice tells owners to contact InkBird for a full refund. Other notices give the company's email address; the links in the table open each one.
Dispose of it as electronic waste
Health Canada's notices say to follow your local electronic waste requirements.
Test again with a method you can trust
A lab-analyzed short-term or long-term kit, or a monitor with an independent result. See radon test kits and radon detectors and monitors.
Act on the new result
EPA recommends fixing a home when the radon level is 4 pCi/L or higher. EPA recommends considering a fix when the level is between 2 and 4 pCi/L. The results interpreter explains a number once you have one.

Read this figure as textHide the text version
| Test | How long it runs | What it is for |
|---|---|---|
| Short-term test | Stays in the home from 2 to 90 days, depending on the device, and for a minimum of 48 hours. | A faster screening result. Because radon levels vary from day to day and season to season, it is less likely than a long-term test to tell you the year-round average. |
| Long-term test | Stays in the home for more than 90 days. | It is more likely than a short-term test to tell you the home's year-round average radon level. |
| Continuous monitor | Logs readings for the whole test, run as a short-term or a long-term test under the same durations in EPA guidance. | Logs a graph over time, so you can see trends and spot anomalies rather than a single number. |

Read this figure as textHide the text version
- Below 2 pCi/L (below 74 Bq/m³)Below both of EPA's guidance levels.
- 2 to 4 pCi/L (74 Bq/m³ to 148 Bq/m³)EPA recommends considering a fix when the level is between 2 and 4 pCi/L.
- At or above 4 pCi/L (148 Bq/m³ and up)EPA recommends fixing a home when the radon level is 4 pCi/L or higher.
How to avoid buying an inaccurate monitor
Health Canada's advice in its 2025 notices is direct: buy only electronic radon monitors that have passed performance testing by C-NRPP, the Canadian certification program it recognizes. In the United States there is no federal list of approved consumer monitors that we can point you to, so C-NRPP's results are the closest public, independent check, with one caveat: C-NRPP tested the Canadian versions, and we have not confirmed that the U.S. versions are identical.
- Look for the model in C-NRPP's latest results, and read what the grades mean on independent accuracy results explained.
- Read the maker's accuracy claim closely. A claim of, say, ±10% after 7 days at 200 names a time and a radon level; a claim with neither is hard to compare.
- Be wary of a device whose screen shows something other than a radon reading. The recalled Air Steward displayed "Coefficient of injury" on its screen.
- Prefer a lab-analyzed kit when a decision is on the line. A short-term or long-term kit is analyzed by a laboratory; see radon test kits.
- Use any monitor for at least three months before you decide on , as C-NRPP advises, and decide on the long-term average.
Questions people ask about radon monitor recalls
Which radon detectors have been recalled?
Health Canada has recalled the Air Steward Portable Radon Monitor (2022), the Spolehli Radon Detector, the Hanchen Home Radon Detector, the Life Basis LSRM001 and LCARM001 detectors and the InkBird Wifi Radon Detector RD-2 (all 2025). It has also issued advisories on the Elifecity Portable Radon Meter (2022) and on the Funny Kitchen HRDM-02 and Boyd Gresham "Toolhome" HRDM-01B detectors (2025).
Why were these radon monitors recalled?
Because they measure radon inaccurately. Health Canada's 2025 notices say its Radiation Protection Bureau sampled and evaluated each device; the 2022 notices say the devices could report a low value regardless of the actual radon level, with errors greater than 90%.
Is my radon detector on the recall list?
Find the brand and model number on the device, its box or its manual, and compare them with the two tables above. Match the model number, not the look of the device: several of these monitors look alike.
What should I do if I own a recalled radon detector?
Stop using it, ask the seller about a refund, dispose of it as electronic waste and test again with a lab-analyzed kit or a monitor with an independent result. Do not rely on any reading it gave you.
Are recalled radon detectors still sold on Amazon in the U.S.?
We could not confirm it either way on 1 October 2026: Amazon.com answered our automated checks with a security page. Every Health Canada notice here concerns units sold on Amazon.ca. Treat any listing that matches a model number on this page as the recalled product.
What does "not approved by C-NRPP" mean?
The device appears on C-NRPP's not-approved list, the devices its 2024/25 report says C-NRPP does not recommend. Some were also recalled by Health Canada; four on the current list carry no Health Canada notice.
Is a radon monitor that is not on C-NRPP's list inaccurate?
Not necessarily. Manufacturers choose whether to enter C-NRPP's comparison, and Health Canada says a device may be on neither list because it is still being tested or reviewed. A missing result is an absence of evidence, which we label "Not included in C-NRPP's 2024/25 evaluation".
Does C-NRPP test radon monitors sold in the United States?
C-NRPP tests units that manufacturers send in, and its comparison is written for Canadian consumers. Where a U.S. model carries the same name, we have not confirmed that it is identical to the unit C-NRPP tested.
Is there a U.S. list of recalled radon monitors?
We have not found a U.S. agency list of consumer radon monitors that failed accuracy testing. The most complete public record we know of is Health Canada's recall notices and C-NRPP's not-approved list, both on this page.
Can a recalled radon monitor be fixed with a firmware update or recalibration?
None of the seven notices offers a fix. Each tells owners to stop using the device, and consumer monitors cannot be professionally calibrated.
Which radon test should I use instead?
A lab-analyzed short-term kit is the quickest way to get a reading you can rely on; a long-term kit or a monitor run for at least three months gives the better picture of your year-round level. See radon test kits and radon detectors and monitors.